Biphasic Defibrillator Rescue 230

  • Description
  • Reviews (0)

Description

BTE (biphasic truncated exponential) discharge type Impedance compensated energy delivery Energy selection from 2 to 230 J directly from the front panel
ECG ‘R’ wave synchro mode or async mode
Sensor SpO2 compatible with BCI
Heart Rate and Spo2 alarm to be setted
Pre and post shock auto printing
Advisory and AED mode in accordance to AAMI DF 39 (optional)
High resolution backlight TFT LCD 5,7” display
Optionals: Pacemaker, SpO2

TECHNICAL SPECIFICATIONS

– BTE (biphasic truncated exponential) discharge type
– Advisory and AED mode according to AAMI DF 39
– Text prompts on the display will guide the operator through the rescue procedures
– Energy selection: from 2 to 230 Joule directly from the front panel
(2-10 step 1 Joule; 10-230 step 5 Joule)
– Charge time: 5 sec for 230 Joule with fully charged battery
– 100 shocks of 230 Joule with fully charged batteries
– Charge/discharge control from the paddles
– ECG “R” wave synchro mode or async mode
– Can be used with disposable paddles (optional)

DISPLAY
– Large (5.7’’) high resolution backlight LCD display (320×240 pixels)
– ECG through the discharge paddles (1 trace) or through 4 leads patient
cable (6 traces): I, II, III, aVR, aVL, aVF, SpO2 trace and values
– Automatic change over when plugging the patient cable
– Filters: 50 Hz/60 Hz, base line, low pass
– Heart rate indicator

ECG MONITORING
Bandwidth: 0.5 to 140 Hz (-3db) with filters off
ECG trace parameters:
– Velocity 5, 12.5, 25 mm/sec
– Gain 5, 10, 20 mm/mV
– heart rate: digital readout on the display from 20 a 260 bpm (± 8% or ± 4 1/min)
Alarm: HR max settable 250 bpm, HR min settable 20 bpm

PRINTER
– Integrated thermal printer for ECG traces and events documentation
hardcopy including SpO2 level in %
– Paper speed: 5, 10, 25 mm/sec
– Paper width: 58 mm
– Operating mode: manual, automatic (10 sec pre and post shock recording)

AED MODE (OPTIONAL)
– Text only (no voice)
– Energy: 150J fixed
– Protocols: AHA 2005 CPR Guidelines
– Analysis time: max. 12 sec
– Shakable rhythms: VF (ventricular fibrillation) with amplitude > 100μV
VT (ventricular tachycardia) with rate >150 bpm

PULSE OXIMETRY (SpO2) (OPTIONAL)
– Oxygen saturation range: 0 to 100%
– Pulse rate range: 18 to 300 pulses per minute
– Accuracy SpO2: 70-100% ± 2 digits for adults using the finger clip sensor
– SpO2 Alarms (min. 65%)

NON-INVASIVE PACEMAKER (OPTIONAL)
– Rectangular wave
– Pulse Time:22 ms
– Pulse Current: MAX 150 mA
– Pulse Voltage: MAX 150 V
– Operation Mode: Fixed, On demand

POWER SUPPLY
– Internal Ni-Cd high capacity batteries with fast charge time
– Charging time (95%):2 hours
– Mains adapter charger:100-230 V,±10% 50/60 Hz

OPERATING CONDITIONS
– Operating temperature: 0 to 55°C
– Storage temperature: -20 to 65°C
– Relative air humidity: 15 to 80%
– Water proof: IPX4

SAFETY FEATURES
– Two hand safety shock release
– ECG input isolation CF class
– Internal discharge after 30 sec of standby after charge

DIMENSIONS: 340x260x130 mm – WEIGHT: 5.5 kg

OPTIONALS
– AED / Advisory mode
– SpO2
– Complete set of pediatric paddles
– Car-light adapter
– Disposable paddles
– Carrying case
– Pacemaker
– Internal defi paddles

Reviews

There are no reviews yet.


Be the first to review “Biphasic Defibrillator Rescue 230”